Model: M0007

门尼粘度是以一个标准的转子,以恒定的转速(一般2转/分),

在密闭室的试样中转动。转子转动所受到的剪切阻力大小与

试样在硫化过程中的粘度变化有关,可通过测力装置显示在

以门尼为单位的刻度盘上,以相同时间间隔读取数值可作出

门尼硫化曲线,当门尼数先降后升,从最低点起上升5个单位

时的时间称门尼焦烧时间,从门尼焦烧点再上升30个单位的

时间称门尼硫化时间。

这款门尼粘度计主要应用于橡胶及其他弹性材料,是作为一种

标准方法测试原材料或化合物的粘度,并可测试硬质橡胶的加

工特征。

应用:

?合成橡胶

?合成塑胶

?合成塑料

产品特点:

? 气动关闭模具

? 定时器:可控制从高压到低压的时间

? 零点重启

?定时器

参考标准:

? ASTM D1646

电气连接:

? 220/240 VAC @ 50 HZ or 110 VAC @ 60 HZ

(可根据客户需求定制)

外形尺寸:

? H: 1,800mm ? W: 560mm ? D: 560mm

? Weight: 165kg

旋转粘度计NDJ-5S使用标准操作规程

旋转粘度计NDJ-5S使用标准操作规程 1.目的 制定旋转粘度计NDJ-5S使用标准操作规程,使操作达到规范化、标准化,确保数据的准确性。 2.范围 本规程适用于上海地学仪器研究所NDJ-5S旋转粘度计的操作。本仪器具有测量灵敏度高。测定结果可靠,使用操作方便,是用来测量牛顿型液体的绝对粘度和非牛顿型液体的表观粘度的仪器。 3.内容 仪器的操作的使用 开机:开机前,将黄色保护盖帽取下,显示屏亮。但电机不工作,预热20min. 准备被测液体,将被测液体置于直径不小于60mm,高度不低于120mm的烧杯或直筒形容器中。 准确地控制被测液体的温度,恒度至25℃±1℃。 仔细调整仪器的水平,检查仪器的水准器气泡是否居中,保证仪器处于水平的工作状态。 参照量程表(表1),选择适配的转子连接头(向右旋装上,向左旋卸下)。估算样品的粘度范围,根据合适的粘度范围选择相应的转子和转速,当估计不出被测液体的大致粘度时,应视为较高粘度,试用由小到大的转子(转子号由高到低)和由慢到快的转速。原则上高粘度的液体选用小转子(转子号高),则转速,低粘度的液体选用大转子(转子号低),快转速。 (表1)NDJ-5S量程表

缓慢调节升降旋钮,调整转子在被被测液体中的高度,直至转子的液体标志(凹槽中部)与液面相平。 参数设定及测定 打开仪器背面的电源开关,进入等待状态,仪器采用中英文显示。 按“▲”或“▼”键选择所需语言模式,按“1#”处, 按“?”或“?”键选择所需转子号,转子号为5种,即“1#、2#、3#、4#及0#“转子。 按“▲”或“▼”键可切换到转速位置。例台光标停在“60转/分”处,按“?”或“?”键可旋转所需的转速。NDJ-5S转速分为9档,分别为转/分、转/分、转/分、3转/分、6转/分、12转/分、30转/分、60转/分及自动档。 当选择好转子和转转速档位后,按“ok/确定“键,转子开始旋转,仪器开始测量,当右边坚条方块显示光标由下向上升至落刻度时,屏幕显示的粘度值即为测量什。测量 时按”开始/停始“键,仪器将会停止测量;如按” 转子号和转速进行测量。 每个试样应测量两次,测量结果取两次测量的算术平均什。两次测量结果之差小于或等于两次测量结果平均什的10%,否则测量第三次。 仪器具有超称报警功能,若测量值大于100%,测量值显示over。为保证测量精度,测量时量程百分比读数应控制在20%-90%之间,能控制在35%-75%之间为较理想值。 在任何状态下,按“开始/停始”键,程序将从起状态开始运行,操作界面回到用户选择工作状态。 每次使用后应旋出转子,及时清洁转子和保护架,转子擦干净后放回到转子架中。即忌用硬物刮、擦转子,以免破坏转子结构。不可把转子留在仪器上进行清洁。 当测不同样品时,应首先清洁(擦干净)转子和转子保护框架,防止由于被测液体相混淆而引起的测量误差。 注意事项 做到下列各点才能测得较精确的粘度:

涂料粘度及其测定知识

涂料粘度及其测定知识 0 前言 粘度是涂料性能中的一个重要指标,对于涂料的储存稳定性,施工性能和成膜性能有很大影响。 例如对于乳胶漆,在贮存过程中涂料的剪切应力ъ>lO dyn/cm2有利于防止沉降,粘度15-30 Pa·s能保证适当的沾漆量;粘度在2.5~5.0 Pa·s保证刷涂性和最佳漆膜性能。在刷涂后如果粘度能够>250 Pa·s 则能很好地控制流挂,因此测定涂料的粘度成为涂料生产和检验中的常规项目。 1 粘度的定义 粘度可以认为是液体对于流动所具有的内部阻力。 动力粘度是指对液体所施加的剪切应力与速度梯度的比值,其国际单位为帕斯卡·秒(Pa·s),习用单位为厘泊(cP)。l cP=1 mPa·s。通过比较在不同剪切速率下粘度的变化。我们可以把流体分为牛顿型流体和非牛顿型流体。在国家标准GB/T 6753.4._l998中将流体的流动类型分为牛顿型流动和不规则流动。 牛顿型流动,当剪切应力与速度梯度比值既不随时间也不随速度梯度方式而改变时,这种材料所呈现的流动类型称为牛顿型流动,当这一比值变化很小时。机械扰动(如搅拌)对粘度的影响可忽略不计,这种材料被称为具有近似牛顿型的流动。一般清漆和低粘度色漆属于这种液体。 不规则流动,当剪切应力与速度梯度比值随时间或随剪切速率而改变

时。这种材料所呈现的流动类型称为不规则流动。 2 涂料粘度的测定方法 涂料粘度的测定方法很多,包括流出杯、斯托默粘度计、落球粘度计、旋转粘度计、毛细管粘度计,锥板粘度计等等。 2.1 涂料粘度测定的国家标准 2.1.1 流出杯法 流出杯是在实验室,生产车间和施工场所最容易获得的涂料粘度测量仪器。由于流量杯容积大,流出孔粗短,因此操作、清洗均较方便,且可以用于不透明的色漆。流量杯粘度计所测定的粘度为运动粘度,即为一定量的试样。在一定温度下从规定直径的孔所流出的时间,以秒表示。这是最常用的涂料粘度测定方法。因为可以在很多场合方便地使用,因此在世界各地得以广泛的应用。 在国家标准中,关于流出杯测涂料粘度的方法标准有GB/T 1723-1993涂料粘度测定法和GB/T6753.4_l988色漆和清漆用流出杯测定流出时间。 在GB/T 1723-1993中使用涂一l杯和涂-4杯。涂一l杯用于测定流出时间不低于20 s的涂料产品。涂一4杯适用于测定流出时间在150 s 以下的涂料。比较两次测定值之差不大于平均值的3%,取两次测定值的平均值作为测定结果。 在GB/T 6753.4一l988中,使用尺寸相似而流出孔径分别为3 mm,4 mm,5 mm,6 mm的4种流出杯,用于测定能准确地判定自流出杯流出孔流出的液流断点的实验物料。对于流出时间超过100 s的实验物

粘度测试注意事项及乌氏粘度计原理

粘度测试注意事项及乌氏粘度计原理 根据其测量原理,为了获得准确可靠的测量数据必须注意以下几点: 一、仪器的性能指标必须满足国家计量检定规程度要求。使用中的仪器要进行周期检定,必要时(仪器使用频繁或处于合格临界状态)要进行中间自查以确定其计量性能合格,系数误差在允许范围内,否则无法获得准确数据。 二、特别注意被测液体的温度。许多用户忽视这一点,认为温度差一点无所谓,我们的实验证明:当温度偏差0.5℃ 时,有些液体粘度值偏差超过5% ,温度偏差对粘 粘度计 度影响很大,温度升高, 粘度下降。所以要特别注意将被测液体的温度恒定在规定的温度点附近,对精确测量最好不要超过0.1℃。 三、测量容器(外筒)的选择。对于双筒旋转粘度计要仔细阅读仪器说明书,不同的转子(内筒)匹配相应的外筒, 否则测量结果会偏差巨大。对于单一圆筒旋转粘度计,原理上要求外筒半径无限大,实际测量时要求外筒即测量容器的内径不低于某一尺寸。例如上海天平仪器厂生产的NDJ-1型旋转粘度计,要求测量用烧杯或直筒形容器直径不小于70mm。实验证明特别在使用一号转子时,若容器内径过小引起较大的测量误差。 四、正确选择转子或调整转速,使示值在20~90格之间。该类仪器采用刻度盘加指针方式读数,其稳定性及读数偏差综合在一起有0.5格,如果读数偏小如5格附近,引起的相对误差在10%以上,如果选择合适的转子或转速使读数在50格,那么其相对误差可降低到1%。如果示值在90格以上,使游丝产生的扭矩过大,容易产生蠕变,损伤游丝,所以一定要正确选择转子和转速。 五、频率修正。对于国产仪器名义频率在50Hz,而我国目前的供电频率也是50 Hz,我们用频率计测试变动性小于0.5%,所以一般测量不需要频率修正。但对于日本和欧美的有些仪器, 名义频率在60Hz, 必须进行频率修正,否则会产生20%的误差,修正公式为: 实际粘度=指示粘度×名义频率÷实际频率 六、转子浸入液体的深度及气泡的影响。旋转粘度计对转子浸入液体的深度有严格要求,必须按照说明书要求*作(有些双筒仪器对测试的液体用量有严格要求,必须用量筒量取)。在转子浸入液体的过程中往往带有气泡,在转子旋转后一段时间大部分会上浮消失,附在转子下部的气泡有时无法消除,气泡的存在会给测量数据带来较大的偏差,所以倾斜缓慢地浸入转子是一个有效的办法。 七、转子的清洗。测量用的转子(包括外筒)要清洁无污物,一般要在测量后及时

SNB-2-J 数显旋转粘度计操作规程

SNB-2-J 数显旋转粘度计操作规程 一、目的 规范SNB-2-J数显旋转粘度计的使用和操作。 二、适用范围 适用于SNB-2-J数显旋转粘度计的使用和操作标准。 三、职责 检验员对本标准的实施负责、保养。主管负责设备操作、保养监督检查。 四、工作原理: 本仪器为数显粘度计,由电机经变速带动转子作恒速旋转。当转子在液体中旋转时,液矩也越小。该作用在转子上的粘性力矩由传感器检测出来,经计算机处理后得出被测液体的粘度。 五、主要技术指标: 测量范围:100mPa.s~200000mPa.s 转子规格:21、27、28、29号转子 转子转速:5、10、20、50 转/分 测量方式:手动、自动 测量精度:±2%(牛顿液体) 控温范围:室温+20℃~200℃ 内胆测量容积:10cc 使用环境条件: (1)环境温度:5℃~35℃; (2)相对湿度:不大于80%; (3)供电电源:电压220V±10V,频率50HZ±10HZ; (4)产品附近无强的电磁场干扰,无剧烈震动,无腐蚀性气体。 六、操作步骤: 1、准备被测样品,称量10g样品放入恒温槽中,设置被测样品的试验温度。 2、仔细调整仪器的水平,检查仪器的水准器气泡是否居中,保证仪器处于水平的工作状态(装上保护架)。 3、参照量程表,选择适配的转子旋入转子连接头(向右旋装上,向左旋卸下)。装卸

转子时,必须用手将连接螺杆微微向上抬起。 4、缓慢调节升降旋钮,调整转子在被测液体中的高度,直至转子的液体标志(凹槽中部)与液面相平。 5、试样在测试温度下充分恒温,以保持示值稳定准确。 6、选择合适测量方式和转速,按“确认”键测量。 七、操作说明 1、首先大约估计被测体液的粘度范围,然后根据下列量程表选择合适的转子和转速。 2、当估计不出被测液体的大致粘度时,应视为较高粘度,试用由小到大的转子(转子号由高到低)和由慢到快的转速。原则上高粘度的液体选用小转子(转子号高),慢转速,低粘度的液体选用大转子(转子号低),快转速。 3、本产品具有自动测量方式。当测量粘度范围不明的样品时,可以先不设置转速,只要选定转子,按“确认”键,仪器会自动开始测量,逐步搜索到合适的转速。测量第二遍及以后各遍时,如果继续在自动方式下测量,则按“复位”键复位,再按“确认”键测量。 4、为保证测量精度,测试时扭矩百分比读数应在10~100%的范围内。 5、如果选择的转速使扭矩值低于10%或高于100%时,就会在粘度值下方出现一个闪烁的问号,此时你需要改变转速或者转子使扭矩值在10~100%之间。 八、注意事项 1、本仪器在出厂前严格调校检验,开机后即可正常工作,请操作者在操作前认真仔细地阅读本仪器说明书,严格按要求操作。 2、仪器电源必须在指定的电压和频率误差范围内测定,否则会影响测量精度。 3、装卸转子时应小心操作,要将仪器下部的连接头轻轻地向上托起后进行拆装,不要用力过大,不要使转子横向受力,以免转子弯曲。连接头和转子连接端面及螺纹处应保持清洁,否则将影响转子的正确连接及转动时的稳定性。 4、装上转子后不得在无液体的状况下“旋转”,以免损坏轴尖和轴承。

NDJ-8S旋转粘度计操作规程.docx

NDJ-8S旋转粘度计标准操作规程 1、目的: 通过制定旋转粘度计标准操作规程,掌握粘度计操作方法,确保测试准确无误。 2、适应范围: 适用于NDJ-8S旋转粘度计。 3、依据:NDJ-8S旋转粘度计说明书。 4、责任人员:实验室操作人员。 5、工作原理: 本仪器为数显粘度计,由电机经变速带动转子作恒速旋转。当转子在液体中旋转时液体会产生作用在转子上的粘度力矩。液体的粘度越大,该粘性力矩也越大;反之液体的粘度越小,该粘性力矩也越小。该作用在转子上的粘性力矩由传感器检测出来,经计算机处理后得出被测液体的粘度。 6、面板操作: 6.1 开启仪器背面电源开关,进入等待选择状态。 6.2 按▲或▼键,来进行当前项的数据修改,按?键进行参数项选择,再按“确定”键即可进入测量界面。 6.3 进入测试界面后按▲或▼键进行调速;在测量状态下按?键,可进行显示温度和显示百分比的转换;在测量状态下按?键,把测量的当前数据传送到电脑,或打印机等是否要进行数据切换;按确定键,即可打印当前所测量的各项数据。 6.4如仪器显示温度的值和实际的温度值有出入,可通过修改TC的值来进行温度的修正,修改后按“复位”即可退出初始状态。

6.5 对于未知样品的粘度测量,首先应估算样品的粘度值。再选择相对应的几组转速、转子组合来进行测量。当估计不出被测流体的大致粘度时,应假定被测样品为较高的粘度;试用由小到大的转子(表面积小)和由慢到快的转速。粘度测量的原则是高粘度的流体选用小的转子(表面积小)、慢转速;低粘度的流体选用大(表面积大)的转子和快的转速进行测试。 6.6转子与转速的组合所对应的粘度范围,可参考下表: NDJ-8S量程表 转子 满量程值mPa.s 1 2 3 4 转速r/min 60 100 500 2000 10000 30 200 1000 4000 20000 12 500 2500 10000 50000 6 1000 5000 20000 100000 3 2000 10000 40000 200000 1.5 4000 20000 80000 400000 0.6 10000 50000 200000 1000000 0.3 20000 100000 400000 2000000 7、操作步骤 7.1 开机:开机前,将黄色保护盖帽取下,显示屏亮,但电机不工作,预热 20min。

NDJ-1旋转粘度计操作规程

SOP/QC(07)016-01 旋转粘度计操作及预防性维护 操作规程 文件类别:操作规程 审批表 江西中兴汉方药业有限公司

目的:制定旋转粘度计操作规程,规范旋转粘度计操作,保证旋转粘度计正常运行。依据:厂家说明书 范围:适用于旋转粘度计操作。 责任:质量控制科QC主任及QC检验员 正文: 1 程序 1.1 仪器与用具 1.1.1 旋转式粘度计 1.1.2 恒温水浴 1.1.3 温度计,分度0.2 ℃ 1.2 操作方法 1.2.1 仪器安装及操作按仪器使用说明进行,并根据供试品的粘度范围和药典在该品种正文 项下的规定,选用是适宜转子和转速。 1.2.2 按各该药品项下的测定温度调整恒温水浴温度。 1.2.3 取供试品置仪器规定的容器中,恒温30 分钟后,依法测定偏转角(α)。关闭马达。 1.2.4 另取供试品同法操作,取二份供试品测定平均值 1.2.5 取2 次测定的平均值按公式计算,即得供试品的动力粘度。 1.2.6 测定时当指针稳定后即应读数,经一定时间旋转后粘度值会逐渐下降。 1.3 记录记录旋转式粘度剂型号,所用转子号数及转速,测定温度等。 1.4 NDJ-1 型旋转式粘度计的标准操作规程 1.4.1 准备被测液体,置于直径不小于70mm的烧杯或直筒形容器中,准确地控制被液体 的温度。 1.4.2 将保护框架装在仪器上(向右旋入装上,向左旋出卸下)。 1.4.3 将选配好的转子旋入连接螺杆(向左旋入装上,向右旋出卸下)。旋转升降钮,使仪 器缓慢地下降,转子逐渐浸入被测液体中,直至转子液面标志与液面平行为止。调整仪器水平,开启电机开关,转动变速旋钮,使所需转速数向上,对准速度指示点,转子在液体中旋转。经过多次旋转,一般为(20~30)s,或按规定时间,待指针趋于稳定可进行读数。按下指针控制杆,使读数回定下来,待指针转至读数窗口时关闭电机(注意:1、不得用力

NDJ-8S旋转粘度计操作规程.docx演示教学

N D J-8S旋转粘度计操作规程.d o c x

NDJ-8S旋转粘度计标准操作规程 1、目的: 通过制定旋转粘度计标准操作规程,掌握粘度计操作方法,确保测试准确无误。 2、适应范围: 适用于NDJ-8S旋转粘度计。 3、依据:NDJ-8S旋转粘度计说明书。 4、责任人员:实验室操作人员。 5、工作原理: 本仪器为数显粘度计,由电机经变速带动转子作恒速旋转。当转子在液体中旋转时液体会产生作用在转子上的粘度力矩。液体的粘度越大,该粘性力矩也越大;反之液体的粘度越小,该粘性力矩也越小。该作用在转子上的粘性力矩由传感器检测出来,经计算机处理后得出被测液体的粘度。 6、面板操作: 6.1 开启仪器背面电源开关,进入等待选择状态。 6.2 按键,来进行当前项的数据修改,按键进行参数项选择,再按“确定”键即可进入测量界面。 6.3 进入测试界面后按 键,把测量的当前数据传送到 电脑,或打印机等是否要进行数据切换;按键,即可打印当前所测量的各项数据。 6.4如仪器显示温度的值和实际的温度值有出入,可通过修改TC的值来进行温度的修正,修改后按“复位”即可退出初始状态。

6.5 对于未知样品的粘度测量,首先应估算样品的粘度值。再选择相对应的几组转速、转子组合来进行测量。当估计不出被测流体的大致粘度时,应假定被测样品为较高的粘度;试用由小到大的转子(表面积小)和由慢到快的转速。粘度测量的原则是高粘度的流体选用小的转子(表面积小)、慢转速;低粘度的流体选用大(表面积大)的转子和快的转速进行测试。 6.6转子与转速的组合所对应的粘度范围,可参考下表: 7、操作步骤 7.1 开机:开机前,将黄色保护盖帽取下,显示屏亮,但电机不工作,预热 20min。

实验二--乌氏粘度计测定聚合物的特性粘度

实验二--乌氏粘度计测定聚合物的特性粘度

实验二乌氏粘度计测定聚合物的特性粘度 一、实验目的 粘度法是测定聚合物分子量的相对方法,此法设备简单,操作方便,且具有较好的精确度,因而在聚合物的生产和研究中得到十分广泛的应用。 通过本实验要求掌握粘度法测定高聚物分子量的基本原理、操作技术和数据处理方法。 二、实验原理 分子量是表征化合物特征的基本参数之一。但高聚物分子量大小不一,参差不齐,一般在103~107之间,所以通常所测高聚物的分子量是平均分子量。测定高聚分子量的方法很多,本实验采用粘度法测定高聚物分子量。 高聚物在稀溶液中的粘度,主要反映了液体在流动时存在着内摩擦。在测高聚物溶液粘度求分子量时,常用到下面一些名词。 如果高聚物分子的分子量愈大,则它与溶剂间的接触表面也愈大,摩擦就大,表现出的特性粘度也大。特性粘度和分子量之间的经验关系式为: 式中,M 为粘均分子量;K为比例常数;alpha是与分子形状有关的经验参数。K和alpha值与温度、聚合物、溶剂性质有关,也和分子量大小有关。K 值受温度的影响较明显,而alpha值主要取决于高分子线团在某温度下,某溶剂中舒展的程度,其数值解与0.5~1 之间。K 与alpha 的数值可通过其他绝对方法确定,例如渗透压法、光散射法等,从粘度法只能测定[η]。 在无限稀释条件下 因此我们获得[η]的方法有二种;一种是以ηsp/C对C 作图,外推到C→0 的截距值;另一种是以lnηr/C对C作图,也外推到C→0 的截距,两根线会合于一点。方程为:

测定粘度的方法主要有毛细管法、转筒法和落球法。在测定高聚物分子的特性粘度时,以毛细管流出发的粘度计最为方便若液体在毛细管粘度计中,因重力作用流出时,可通过泊肃叶公式计算粘度。 (m=1)。 对于某一只指定的粘度计而言,(4)可以写成下式 省略忽略相关值,可写成: 式中,t 为溶液的流出时间;t0为纯溶剂的流出时间。 可以通过溶剂和溶液在毛细管中的流出时间,从(6)式求得ηr,再由图求得[η]。 三、实验主要仪器设备和材料 主要仪器:恒温玻璃水浴(包括电加热器、电动搅拌器、温度计、感 温元件和温度控制仪)、三管乌式粘度计、秒表、洗 耳球、 250ml 三角烧瓶、20ml移液管、40 ml砂芯 漏斗 主要原料:溶剂(分析纯)和聚合物自选 四、实验方法、步骤及结果测试 1. 试样准备: 按溶剂选择原则选择待测高聚物的溶剂。从手册查所选高聚物/溶剂对在特定温度下Mark-Houwink方程中的K和α值。 预先在容量瓶内配制精确体积的溶液。浓度选择要使溶液和纯溶剂流经乌氏粘度计上两刻度线之间C球的时间比约为1.2~2.0。 2. 温度调节:

USP34-NF29 911 粘度 中文翻译

<911> VISCOSITY Viscosity is a property of liquids that is closely related to the resistance to flow. It is defined in terms of the force required to move one plane surface continuously past another under specified steady-state conditions when the space between is filled by the liquid in question. It is defined as the shear stress divided by the rate of shear strain. The basic unit is the poise; however, viscosities commonly encountered represent fractions of the poise, so that the centipoise (1 poise = 100 centipoises) proves to be the more convenient unit. The specifying of temperature is important because viscosity changes with temperature; in general, viscosity decreases as temperature is raised. While on the absolute scale viscosity is measured in poises or centipoises, for convenience the kinematic scale, in which the units are stokes and centistokes (1 stoke = 100 centistokes) commonly is used. To obtain the kinematic viscosity from the absolute viscosity, the latter is divided by the density of the liquid at the same temperature, i.e., kinematic viscosity = (absolute viscosity)/(density). The sizes of the units are such that viscosities in the ordinary ranges are conveniently expressed in centistokes. The approximate viscosity in centistokes at room temperature of ether is 0.2; of water, 1; of kerosene, 2.5; of mineral oil, 20 to 70; and of honey, 10,000. 粘度是液体的属性之一,它与流动阻力紧密相关。在规定的条件下,待测液体充满平面间的空隙,在不断转动过程中,所产生的力定义为粘度。该粘度等于剪切应力除以剪切应变率。基本单位是泊;但是经常遇到泊的分数表示的粘度,因此厘泊(1泊=100厘泊)是更常用的单位。由于粘度随温度变化明显,需指明温度。通常情况下,粘度随温度的升高而减小。尽管粘度的测量是以绝对粘度形式表示,其常用单位是泊或厘泊,但为方便期间常常得到的是运动粘度,常用单位是司和厘司(1司=100厘司)。为了从绝对粘度得到运动粘度,绝对粘度需要除以同温度下的液体密度。如,运动粘度=绝对粘度/密度。单位的大小便于表示常规范围内的粘度为厘司。在室温情况下,以厘司估计的粘度,醚为0.2;水为1;煤油为2.5;矿物油为20~70;蜂蜜为10,000。 Absolute viscosity can be measured directly if accurate dimensions of the measuring instruments are known, but it is more common practice to calibrate the instrument with a liquid of known viscosity and to determine the viscosity of the unknown fluid by comparison with that of the known. 如果已知测量仪器的准确尺寸,可以直接测得绝对粘度。但是更常用的做法是使用已知粘度的液体校准仪器,通过与已知粘度的液体相对比,测定未知粘度的液体。 Many substances, such as the gums employed in pharmacy, have variable viscosity, and most of them are less resistant to flow at higher flow rates. In such cases, a given set of conditions is selected for measurement, and the measurement obtained is considered to be an apparent viscosity. Since a change in the conditions of measurement would yield a different value for the apparent viscosity of such substances, the instrument dimensions and conditions for measurement must be closely adhered to by the operator. 许多物质,如制药中用到的胶,具有可变粘度。它们中的大多数在高流速的情况下,流动阻力较低。在这种情况下,为了测量粘度选定一个特定条件,得到的测量值被认为是表观粘度。因此测量条件的改变会造成这类物质的表观粘度值的不同。操作者必须严格遵守测量时的仪器尺寸和条件。 Measurement of Viscosity— The usual method for measurement of viscosity involves the determination of the time required for a given volume of liquid to flow through a capillary. Many capillary-tube viscosimeters have been devised, but Ostwald and Ubbelohde viscosimeters are among the most frequently used. Several types are described, with directions for their use, by the American Society for Testing and Materials (ASTM, D-445). The viscosity of oils is expressed on arbitrary scales that vary from one country to another, there being several

旋转粘度计实用标准操作规程

实用文案 旋转粘度计使用说明书

一、概述 NDJ-1型旋转粘度计是根据上海市企业标准Q/YXYY 20-2000《NDJ-1型旋转式粘度计》规定的技术要求设计和制造的,它可广泛应用于对油脂、油漆、塑料、食品、药物、胶粘剂等各种流体粘度的测量。 二、主要技术指标及参数 1、测量范围:(10~100000)mPa·s;

2、测量误差:±5%(F·S); 3、测量转子:1号、2号、3号、4号转子; 4、转子转速:6转/分、12转/分、30转/分、60转/分 5、供电电源:AC220V±10% 50Hz±10%; 6、外形尺寸:300㎜×300㎜×450mm; 7、净重: 1.5kg(不包括支架)。 三、仪器结构和安装 (一)仪器结构 1、结构原理 结构原理图见图1所示。 图1 ⑴同步电机以稳定的转速旋转,连接刻度圆盘,再通过游丝和转轴带动转子旋转。如果转子未受到液体的阻力,则游丝、指针与刻度盘同速旋转,指针在刻度圆盘上指出的读数为“0”。而当转子受到液体的粘滞阻力,则游丝产生扭矩,与粘滞阻力抗衡,最后达到平衡,这时与游丝连接的指针在刻度圆盘上指示一定的读数(即游丝的扭转角)。 将读数乘上特定的系数即得到液体的粘度(mpa·s)。

⑵利用齿轮系统及离合器进行变速,由专用旋转旋钮操作,分四档转速,可以根据测定需要选择。 ⑶按仪器不同规格附有1至4号四种转子,可根据被测液体粘度的高低随同转速配合使用。 ⑷为使读数精确,仪器装有指针固定控制装置(指针控制杆)。当转速较快时(30转/分,60 转/分),无法在旋转时读数,这时可以按下指针控制杆,使指针固定下来,便于读取准确的读数。 ⑸保护架是为了稳定测量和保护转子而专门设计的。使用保护架进行测量能取得较稳定的测量结果。 ⑹整套仪器配有固定支架和升降机构,一般在实验室中进行小量和定温测定时应固定使用。另外,仪器也可以脱离固定支架和升降机构手提使用。 2、整体结构 ⑴机头的结构示意图见图2所示。

粘度测试注意事项及乌氏粘度计原理讲课稿

粘度测试注意事项及乌氏粘度计原理

粘度测试注意事项及乌氏粘度计原理 根据其测量原理,为了获得准确可靠的测量数据必须注意以下几点: 一、仪器的性能指标必须满足国家计量检定规程度要求。使用中的仪器要进行 周期检定,必要时(仪器使用频繁或处于合格临界状态)要进行中间自查以确定其计量性能合格,系数误差在允许范围内,否则无法获得准确数据。 二、特别注意被测液体的温度。许多用户忽视这一点,认为温度差一点无所谓,我们的实验证明:当温度偏差0.5℃ 时,有些液体粘度值偏差超过5% ,温度偏差对粘 粘度计 度影响很大,温度升高, 粘度下降。所以要特别注意将被测液体的温度恒定在规定的温度点附近,对精确测量最好不要超过0.1℃。 三、测量容器(外筒)的选择。对于双筒旋转粘度计要仔细阅读仪器说明书,不同的转子(内筒)匹配相应的外筒, 否则测量结果会偏差巨大。对于单一圆筒旋转粘度计,原理上要求外筒半径无限大,实际测量时要求外筒即测量容器的内径不低于某一 尺寸。例如上海天平仪器厂生产的NDJ-1型旋转粘度计,要求测量用烧杯或直筒形容器直径不小于70mm。实验证明特别在使用一号转子时,若容器内径过小引起较大的 测量误差。 四、正确选择转子或调整转速,使示值在20~90格之间。该类仪器采用刻度盘加指针方式读数,其稳定性及读数偏差综合在一起有0.5格,如果读数偏小如5格附近,引起的相对误差在10%以上,如果选择合适的转子或转速使读数在50格,那么其相对误差可降低到1%。如果示值在90格以上,使游丝产生的扭矩过大,容易产生蠕变,损伤游丝,所以一定要正确选择转子和转速。 五、频率修正。对于国产仪器名义频率在50Hz,而我国目前的供电频率也是50 Hz,我们用频率计测试变动性小于0.5%,所以一般测量不需要频率修正。但对于日 本和欧美的有些仪器, 名义频率在60Hz, 必须进行频率修正,否则会产生20%的误差,修正公式为: 实际粘度=指示粘度×名义频率÷实际频率 六、转子浸入液体的深度及气泡的影响。旋转粘度计对转子浸入液体的深度有严格要求,必须按照说明书要求*作(有些双筒仪器对测试的液体用量有严格要求,必须用量筒量取)。在转子浸入液体的过程中往往带有气泡,在转子旋转后一段时间大

旋转式粘度计标准操作规程

XXXXXXXX仪器设备标准操作规程 1 目的:建立NDJ-1型旋转式粘度计标准操作规程。 2 范围:NDJ-1型旋转式粘度计。 3 责任:化验室操作者。 4 内容: 4.1 首先将保护架右旋装上仪器下端。 4.2 准备被测液体选择直径≥70mm的圆柱形容器。 4.3 按测量要求的精度,准确控制被测液体的温度。 4.4 估计被测液体粘度值的大致范围参考量程表选配好转子号,并将转子左旋入连接螺杆。然后旋转升降钮,使仪器慢慢地下降,逐渐浸入测量液体中,直至转子液面标志和液面平为止。按下指针锁定杆,开启电源开关,放开指针锁定杆,使转子在液体中旋转,一般20~30秒的运转,待指针趋于稳定。按下指针锁定杆,使计数固定下来,再关闭电机,使指针停在读数窗范围内,应继续按住指针锁定杆,反复开启和关闭电机,使指针处于读数窗内。如选用6和12转/分时,可直接读取读数,不需锁定指针。 4.5 读数指针一般在30~90分度之间为佳,否则可变换转子号和转速达到这个要求。 4.6 当不能预知被测液体粘度值时,应试用由小到大的转子和转动变速钮,由慢到快的变换转速顺序来测定粘度值。

5.1 测定时指针在刻度盘上的指示值乘上系数表中的系数值,得到测量液的绝对粘度值。 即:η(mpa.s)=k×α K=系数 α=指针读数值 η=粘度值 5.2 频率误差修正 实际粘度=指示粘度×名义频率/实际频率 系数表量程表 6 维护及注意事项 6.1 本仪器使用于10~30℃。 6.2 测定时必须装上保护架,否则测定粘度将因容器的直径变化而变化。 6.3 转子在安装和拆除时应将转子连接螺杆向上抬起,避免转轴轴尖碰损,浸入液体时避免气泡粘附于转子上。 6.4 仪器不得侧放或倒放,特别是装上转子时。 6.5 装上转子后,不要在未浸入液体的情况下,开机旋转。 6.6 每次使用完毕后小心拆下转子,彻底清洗,用吹风机吹干,垂直安放在转子架上。

旋转式粘度计操作规程

旋转式粘度计操作规程 目的: 制订NDJ-79型旋转式粘度计标准操作规程。 适用范围: NDJ-79型旋转粘度计操作。 责任: 检验室操作人员按本规程操作,检验室主任监督本规程执行。 规程: 1.操作方法(第Ⅲ单元测定) 1.1根据待测样品的黏度范围选择转筒。 附:每个转筒的因子与测量范围。 因子范围 0.1 1~10mpa 0.2 2~20mpa 0.4 4~40mpa 0.5 5~50mpa 1.2零点校正:打开电源开关在电动机空载旋转情况下,将本机左端的调零螺丝轻轻旋入此时指针应指在零点。测试时的零点校正应在空载下反复三次,确认零位无误才算调零结束。 1.3将u形弹簧的两端伸至转筒内,使挂钩露在筒外。 1.4将被测液体小心倒入测试容器,直至液面达到锥形面下部边缘,将转筒插入液体直到完全浸没为止。 1.5将测试容器安放在仪器托架上,并将转筒挂钩悬挂于仪器左旋滚花螺目的挂钩上。 1.6开动电机,此时可得测试器在托架上前后左右轻微移动,使转筒对准中心(此时应没有振动的声音)。 1.7当指针稳定后即可读数,如果读数小于10格,应调换大一号的转筒。 1.8连续测定2次,每次测定值与平均值之差不得超过平均值的±3%,否则应作第三次测定。 1.9测定完毕,取下测试容器,清洗转筒与测试容器,并放回原处。 2.0将调零螺丝退出,关掉电源,并填写“仪器使用记录”。 2.注意事项:

2.1电源线必须先插入仪器后,按通电源,严格按照操作步骤进行使用。 2.2仪器内零件不得随意拆动,游丝严禁拆装。 2.3使用前及使用后,都应将转筒及测定器内壁洗净擦干,以保证仪器测量的精度。 2.4仪器存放处应经常保持干燥。 RCZ-5A智能药物溶出仪标准操作规程 目的:规范溶出度测定仪的操作。 适用范围:检验室RCZ-5A智能药物溶出仪。 责任:检验室操作人员对该规程的实施负责,检验室主任对本规程的有效执行承担监督检查责任。 程序: 1.操作方法: 1.1给水浴箱注纯化水至水线标志。 1.2把电源线甩出端的圆形插头接在有地线的AC220V单相电网的电源插座中。 1.3开机:按下仪器面板左端的电源开关,水泵启动,水浴箱中的水开始流动,转速窗中,“预置”指示灯亮,数码显示计算机初始设定的预置转速100r/min;时钟窗中,“常规”指示灯亮,数码显示常规取样预置时间1小时;温度窗中,“水温”指示灯亮,数码显示水浴箱中水的实际温度。 1.4温度控制:观察显示器的实际水温,若与要求的控制温度相差甚远,可适量更换水浴箱的水,以缩短达到恒温状态所需的时间。 1.5控温与预置:按一下“选择”键,“控温”和“预置”指示灯亮,“水温”指示灯灭,表示仪器进入自动控温状态,同时也进入了温度预置状态,数码管变为显示本机内部设定的初始预置温度37℃。用户可根据工作需要,以及对于水浴温度与杯内温度之差的具体经验,按动增温键“∧”和减温键“∨”,调整预置温度。 1.6控温与水温:再按一次“选择”键,“控温”指示灯仍亮,“预置”指

乌氏粘度计的主要模块介绍

现如今,科技的进步带动了各行业的发展,乌氏粘度计就是其中的产物之一。接下来,我们就根据具体的介绍,来分析一下主要模块是怎么样的吧。 一、HCT系列高精度恒温浴槽 1、浴槽采用双仓体原理设计和独特的循环方式,保证了整个测量仓内非常小的温度梯度和很高的温度精度,温度均匀度达到≤0.01℃(与CR系列流经式制冷器配合); 2、双仓设计,测量仓内的液面高度可不受浴液的蒸发而长时间维持恒定,保证了测量过程中环境的一致性; 3、安全性:具有高、低浴位报警功能,自动切断加热保护功能; 4、温控部分和恒温浴槽槽体可轻松分离,非常方便清洗浴槽混匀区; 5、温控器采用原装进口,温控精度高且具有自适应PID调节功能; 6、双层透明视窗及独特的照明设计,可以非常直接对浴槽内部及测试过程进行观察; 7、内胆采用无缝全焊接工艺,材料为不锈钢,保证了长时间使用的可靠性; 8、可满足1-6个自动或手动测量。 二、IV粘度测量单元 1、IV粘度测量单元与IV-MC主控制器只需一根UBS数据线的连接;

2、传感器:采用智能近红外光电传感器,保证测量时间可以精确到毫秒级;材料:采用不锈钢铠装,耐腐蚀及寿命长;特殊的检测方式及采用智能近红外光电传感器,可满足测量各种不同颜色的样品,包括纯黑母粒; 3、精准及稳定的定位,使测量间距精确到ISO规定的40mm,符合标准要求; 4、材质:不锈钢; 5、外形尺寸(L*W*H):115×90×480mm。 三、IV-MC主控制器及V1.0操作系统 1、IV-MC主控制器是自动粘度测试仪系统的中央模块,是各个测量单元和PC的连接中枢,控制系统最多能与6个IV测量单元连接,而且与每个测量单元只需一根数据线的连接; 2、IV-MC主控制器与IV测量单元及PC端的连接方式:USB; 3、软件支持Win7和win8及XP系统; 4、最多可控制6个IV测量单元同时工作; 5、权限管理功能; 6、测量结果储存及打印功能;

乌氏粘度计清洗模块详情介绍

乌氏粘度计随着不断地改革创新,已经告别了过去耗材的时代,更加的方便。尤其是清洗模块的诞生,就显得越发简单。下面,我们就来看看其具体是怎么样的吧。 一、产品特点 1、具有在线自动排液、自动清洗、自动干燥、自动浸泡等强大功能。 2、自动在线清洗模块,大大减少了实验人员接触有毒化学试剂的机会,很大程度保证了实验人员的人身安全。 3、采用高压脉冲往复式的自动清洗方式,再难洗的聚合物都可以洗的干干净净。 4、独特的干燥方式,在保证干燥效果及效率的前提下,可采用更环保、更安全的干燥试剂进行干燥。 5、废液排放管路均采用耐强酸强碱的聚四氟或氟橡胶材料。 6、智能化废液处理提醒功能,防止废液溢出。 7、废液分类收集功能,可避免不良化学反应发生的同时,有利于废液的处

理及回收再利用。 8、工作效率:相对于人工手动清洗,工作效率提升了一倍以上。 9、可同时控制最多8个清洗模块工作,并具有分级权限管理功能。具有排液速度、清洗速度、管内清洗次数、干燥时间、浸泡时间等参数都可根据不同实验可自行设定功能。 10、软件操作界面有中文、英文版本,并且具有兼容性,可用于其他厂家的半自动粘度仪。 二、技术参数 1、试剂耗量:<13ml(单次)。 2、通讯接口:RS232。 3、额定功率:100W。 4、设备尺寸:200*200*450mm(长*宽*高)。 5、电源:交流220V/50Hz。 综上所述,就是乌氏粘度计清洗模块的初步讲解,要是感兴趣的小伙伴,可以自行咨询,或者去一些专业生产的厂家进行实地考察。 卓祥科技的团队专注于自动乌式粘度分析行业至少七年以上,且一直拥有几

十位专业的高分子材料研发/生产/实验人员的鼎力支持。同时也一直专注于研发高分子材料等领域的采用粘度分析仪器,设计灵感凝聚了几十家高分子材料生产商的实验人员和科研院所研发人员的智慧。应不同市场所需,公司已先后成功研发推出了AVM系列全自动粘度仪、IV6000系列全自动乌式粘度仪等多种系列产品。

ANDLF-Tech-QI001-01V01粘度检测方法(斯托默粘度计法)

All rights of multiplication are reserved to ANDLF page 1 of 2 They may only be passed on with prior consent of the releasing position. 阿克苏诺贝尔装饰涂料(廊坊)有限公司 Akzo Nobel Decorative Coatings (Langfang) Co., Ltd. 粘度检测方法(斯托默粘度计法) Work Instruction

All rights of multiplication are reserved to ANDLF page 2 of 2 They may only be passed on with prior consent of the releasing position. 1 目的 为规范粘度(斯托默粘度)的测定方法,特制定本作业指导书。 2 适用范围 本作业指导书适用于阿克苏诺贝尔装饰涂料(廊坊)有限公司所有斯托默粘度的测定。 3 检测依据 GB/T 9269-1988 建筑涂料粘度的测定 斯托默粘度计法 4 检测方法 4.1 将涂料充分搅匀移入500毫升漆罐中,使涂料液面离盖约19毫米。 4.2 调节涂料和粘度计的温度至待测样要求温度。 4.3 按下升降杆,将转子浸入涂料中,使液面刚好达到转子轴的标记处。 4.4 按“Read ”键,待读数稳定后,即为待测样品的粘度。 4.5 重复测定三次,取平均值即为检测结果 5 相关文件和记录 5.1 相关文件 ANDLF-Tech-QP001 《标准检测方法》 5.2 相关记录 无 6 文件修改记录